Page 16 DataLoggerXP User’s Manual
Data
Data is copied to the location specied in the Options Dialog. The user must provide the sheet,
row, and column reference. If the reference is invalid, the data will be copied to the default active
cell when the template is opened (usually column A of row 1 in the rst worksheet in the Excel
workbook). When the data is copied, it will overwrite any existing cell contents, but cell formatting
will not be modied.
Default Template File
It is possible to congure DataLoggerXP to always use a specic template le when saving data in
Excel workbook format. By checking the Use Excel template by default checkbox in the Options
Dialog and selecting the Excel template le to use, when saving data, the system will automati-
cally open this template le to create the workbook. The Use Excel template by default option
also changes the default save type from a Comma Separated Text File to an Excel Workbook.
Service and Support
Troubleshooting
This section provides possible solutions to common problems encountered when working with
DataLoggerXP and the XP2i in data logging mode. For additional troubleshooting support, please
see the troubleshooting section in the XP2i operation manual.
Problem: Gauge display is constantly ashing
Solution: This condition indicates the gauge’s onboard program is corrupted. This is typically
caused by a failure during a mode switch, and be recovered by using the File/Repair Gauge
menu option. If this process fails, contact customer service for support.
Problem: The XP2i will not connect with DataloggerXP
Solution: Make sure the gauge is on (and not displaying d o”).