Specications 3
APM CPF Series Operation Manual
Specifications
Modules
Module
Range
0 - 30%
Gauge Full Scale
30 - 110%
Gauge Full Scale
psi ± (% Full Scale) ± (% of Reading)
psi
30 0.0075% 0.025%
100 0.0075% 0.025%
300 0.0075% 0.025%
1000 0.015% 0.05%
3000 0.015% 0.05%
10000 0.03% 0.1%
15000 0.03% 0.1%
bar ± (% Full Scale) ± (% of Reading)
bar
3 0.0075% 0.025%
10 0.0075% 0.025%
30 0.0075% 0.025%
100 0.015% 0.05%
300 0.015% 0.05%
700 0.03% 0.1%
1000 0.03% 0.1%
kPa / MPa ± (% Full Scale) ± (% of Reading)
kPa /
MPa
300 0.0075% 0.025%
1 0.0075% 0.025%
3 0.0075% 0.025%
10 0.015% 0.05%
30 0.015% 0.05%
70 0.03% 0.1%
100 0.03% 0.1%
Accuracy (Gauge)
0 to 30% of Full Scale . . . . . . . . . . . . ±(0.0075% of Full Scale), ±(0.015% of Full Scale), or ±(0.03% of Full Scale)
30 to 110% of Full Scale . . . . . . . . . .±(0.025% of Reading), ±(0.05% of Reading), or ±(0.1% of Reading)
Vacuum . . . . . . . . . . . . . . . . . . . . . . . . . For 100 psi / 10 bar / 1 MPa and lower:
±(0.06% of Full Scale
*
)
For 300 psi / 30 bar / 3 MPa:
±(0.06% of Full Scale
*
) ±1 LSD
*
Full Scale = -14.5 psi / -1.0 bar / -100.0 kPa.
Accuracy specications include all eects of linearity, hysteresis, repeatability, temperature, and stability for one year.
Note: Exposure to environmental extremes of temperature, shock, and/or vibration may warrant a more frequent
recertication period.
APM modules must be exercised and re-zeroed whenever exposed to signicant changes in environmental conditions to achieve
these specications. To exercise a module, cycle the module between zero (ambient barometric pressure) and the pressure of inter-
est.
A properly exercised module will return to a zero reading (or return to the same ambient barometric reading).
!
CAUTION: Advanced Pressure Modules (APM) are not recommended for continuous use at high vacuum.