Specications 5
XP2i-DP Operation Manual
—
PRESSURE RANGES, DISPLAY SCALES, AND RESOLUTION
psi bar kPa MPa Overpressure (+/– ports) psi
in
H
2
0 inHg mmHg mmH
2
0 kg/cm
2
bar mbar kPa MPa
15PSI 1BAR 100KPA 0.1MPA 6.5x / 3.0 x 0.001 0.01 0.001 0.01 1 0.0001 0.0001 0.1 0.01 0.00001
100PSI 7BAR 700KPA 0.7MPA 2.0 x / 2.0 x 0.01 0.1 0.01 0.1 1 0.0001 0.0001 0.1 0.01 0.00001
•
Unneeded pressure units may be disabled via the RS-232 connector using CongXP software.
•
An XP2i-DP will indicate pressure up to 10% above its stated pressure rating. Above 110%, the XP2i-DP display will ash, indicating that the applied pressure
exceeds the calibrated pressure range. If the calibrated pressure range is exceeded, the pressure displayed may not be accurate.
•
Overpressure is the maximum pressure the gauge can withstand without damage. The gauge will not indicate pressure up to this value.
•
kPa and MPa models can display pressure in kPa, MPa, and bar (and/or mbar) only. psi and bar models can display all available units.