Calibrating Gauges 16
FastCalXP Operation Manual
TEST PROFILES
Test pro les are the standardized series of pressure levels used for the calibration of each model. Thus, there is a pro le for As Received calibrations and a
pro le for As Left calibrations associated with each model in the database.
To Select a Test Pro le
Test pro les can be speci ed within the New Model Wizard (accessed via the New Gauge Wizard).
1 Use the AR test pro le and AL test pro
le pull-down lists to select a test pro le for the As Received calibration and a test pro le for the As Left calibration.
2 If the desired pro le is not listed, click the Create New Pro le button to open the Pro les editor.
Refer to the discussion of Test Pro les in the Managing the Database section for help with the Pro les edit
or.
To Edit a Pro le or Add a New Pro le Outside of the New Model Wizard
Choose Manage Database from the Tools menu and click on the Pro les button to open the Pro les editor.
To change the pro les that are associated with a model
1 Choose Manage Database from the Tools menu and click on the Models button to open the Models editor.
2 Select the desired model and choose the desired pro
les fr
om the pull-down lists for AR test pro le and AL test pro le.
3 If a desired pro le is not list
ed, click the Create New Pro le button to open the Pro les editor.
When the pro le has been created, return to the Models editor and select the pro le from the pull-down list.