Settings and Options 28
FastCalXP Operation Manual
Save Certi cates To
The calibration certi cates created by FastCalXP will be saved in the directory speci ed in Save certi cates to. This directory is set the rst time the software is
run after installation or when a new database is created and can be changed here using the Browse button.
Certi cate Naming Template
The Certi cate naming template is used by FastCalXP to name the calibration certi cates. In general, certi cates are numbered with the Next certi cate
number incrementing automatically each time a certi cat
e is created. Use number signs (#’s) in the Certi cate naming template to de ne the number of digits
to be used.
Calibration Representative and Quality Representative
The Calibration representative and the Quality representative must be listed users and can be selected using the pull-down lists. In the certi cate template
provided with the software these names will be listed, with signature lines, on the rst page of each certi cate.
Calibration Settings
The Calibration Settings de ne the default Calibration Method.
Actual
Uses the pressure reading from the reference at the moment the Next Point button is clicked.
Peak
Automatically advances to the Next Point when a peak is detected.
Head Correction
If the gauge and the reference do not sit at the same height, a Head correction can be applied. Enter a correction to be added to the reference pressure, or use
the Correction calculator to calculate a correction given the speci c conditions of the calibration. Refer to the discussion of the Head Correction Calculator
on page 32.