Crystal FastCalXP Calibration Software Manual de usuario Pagina 20

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 59
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 19
Calibrating Gauges 18
FastCalXP Operation Manual
To Change the Number of Exercise Cycles Associated with a Pro le
1 choose Manage Database from the Tools menu and click on the Pro les button.
2 Select the pro le of int
erest and click on the Edit button.
3 Enter the desired number of exercise cycles in the Number of exercise cycles text box.
Matching the Target
Two di erent approaches can be taken while stepping through a calibration pro le: matching the gauge to the Target Pressure or matching the reference to
the Target Pressure.
During a calibration, the Target Pressure is the next pro le value and the Reference and Gauge pressures are indicated below. The Gauge value is, by default, the
same value as the Target Pressure. However, the value can be changed by the user to match what is read from the gauge.
To Match the Gauge to the Target Pressure
1 Read the gauge while turning the handle of the GaugeCalHP.
2 Stop when the gauge reaches the Target Pressure value.
Ma
tching the gauge allows for much faster calibration since interpolation is not required to read the gauge and the Gauge value does not have to be
entered.
Vista de pagina 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 58 59

Comentarios a estos manuales

Sin comentarios