Operation and Maintenance 8
Type T Pump Operation Manual
Operation and Maintenance

GENERATING PRESSURE
Note: It may be necessary to prime the pump, particularly if the pump has been inactive for an extended period of time. Refer to Priming on page 8
for instructions on priming the pump.
1
Check that the vent ll plug is fully open by turning it counter-clockwise.
2 Open the vent valve.
Open the vent plug… then open the vent valve.
3 Set the valve rod selector to control your desired uid output volume.
!
CAUTION: Only move the valve rod selector while the pump handle is stationary, in the up position.
Note: With the valve rod in the out position, uid ow will be at maximum and the maximum achievable pressure will be 1000 psi. To exceed 1000 psi, push
the valve rod in.
Low Pressure/
High Volume
High Pressure/
Low Volume