Type T Pump Operation Manual
7 Examine the port for air bubbles which may adhere to the threads or parts.
Note: If air bubbles are present, repeat steps 2 through 4 until no more air bubbles appear.
8 Replace the check valve plug and tighten partially.
9 If replacing the low pressure/high volume check valve plug... Tighten the check valve plug.
The procedure for the low-pressure/high volume side is now complete.
10 If replacing the high pressure/low volume check valve plug... Use slight pressure on the hand lever to push system uid past the high
pressure/low volume check valve plug to ensure complete purging of the valve port.
11
Tighten the high pressure/low volume check valve plug.
12 Operate the pump handle until pumping becomes dicult.
13 Raise the pump handle to the top of its stoke.
14 Push in the valve rod selector to the high pressure/low volume position.
15 Resume pumping to verify the pump is operating properly.
Pump Does Not Hold Pressure
X
Problem: The pump operates but the pressure declines beyond normal thermodynamic eects.
X
Solution: One of two o-rings may be leaking.
As the pressure drops, observe the hand lever. If the hand lever rises, the discharge check valve is leaking and its o-ring should be replaced.